Shinwa Chemical Industries Ltd.
LC/GC Search / Application Note

Shinwa Chemical Industries
Application Search

Search result
"日本薬局方" Filtered 19 Item.

Graph

GC

Samples
Methyl iodide , iso-Propyl iodide , n-Octane , o-Xylene
Applications
Analysis of hypromellose using Shinwasorb
Column
Silicone DC-200 10% Shinwasorb-S 100/120
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
90℃
Carrier gas
38 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Carbon sulfide , Hydrogen sulfide , Carbon disulfide , Methyl mercaptan , Ethyl mercaptane , Dimethyl sulfide
Applications
Measurements of offensive odor substances using Shinwasorb -3-
Column
1,2,3-TCEP 25% Shinwasorb-S 80/100 ST
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
70℃
Carrier gas
50 mL/min N₂
Injection port temp.
180℃
Detector temp.
180℃
Detector
FPD
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Chloroform , Ethanol , Naphthalene , Amantadine
Applications
Measurements of amantadine hydrochloride using Shinwasorb
Column
Apiezone L+KOH 2+1% Shinwasorb-S 80/100
Column Size
2.1 m x 3.2 mm I.D. Glass
Column temp.
125℃(5 min hold)~150℃(5℃/min)~150℃ (10 min hold)
Carrier gas
50 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
200 ppm in hexane
Applications
Measurements of eucalyptus oil using Shinwasorb
Column
Thermon-3000 5% Shinwasorb-S 80/100
Column Size
5.1 m x 3.2 mm I.D. Glass
Column temp.
120℃ & 130℃
Carrier gas
40 mL/min N₂
Injection port temp.
200℃
Detector temp.
200℃
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Isopropyl iodide , Methylcyclohexane , o-xylene
Applications
Analysis of hydroxypropylcellulose (JP)
Column
ULBON HR-1
Column Size
0.53 mm I.D. x 30 m df = 3.0 µm
Column temp.
40℃ (3 min)~10℃/min~100℃~50℃/min~250℃ (3 min)
Carrier gas
52 cm/sec He
Injection port temp.
180°C
Detector temp.
280°C
Detector
FID
Injection volume
2.0 μL Split rate: 1:50

View detail


Graph

GC

Samples
Ethylene glycol , Diethylene glycol
Applications
Analysis of macrogol 400 using Shinwasorb
Column
Sorbitol 3% Shinwasorb-S 80/100
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
160℃
Carrier gas
47 mL/min N2
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Cyclohexane , N,N-dimethylaniline , Naphthalene ,
Applications
Analysis of ampicillin hydrate using Shinwasorb
Column
Silicone OV-17 1.5% Shinwasorb-S 60/80
Column Size
2.1 m x 2.6 mm I.D. Glass
Column temp.
105℃
Carrier gas
20 mL/min He(adjust so that the retention time of N,N-dimethylaniline is about 5 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Cyclohexane , Ethyl acetate , 2-propanol ,
Applications
Analysis of lenampicillin hydrochloride using Shinwasorb
Column
Quadrol 5% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
100℃
Carrier gas
70 mL/min N₂ (Adjust so that the retention time of cyclohexane is about 1 minute.)
Injection port temp.
160°C
Detector temp.
160°C
Detector
FID
Injection volume
4.0 μL

View detail


Graph

GC

Samples
Enflurane , 2-propanol ,
Applications
Analysis of enflurane using Shinwasorb
Column
DEGS 10% Shinwasorb-S 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
80℃
Carrier gas
32 mL/min He (Adjust so that the retention time of enflurane is about 3 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
5.0 μL

View detail


Graph

GC

Samples
Isoflurane , Ethyl acetate
Applications
Analysis of isoflurane using Shinwasorb
Column
Igepal CO-880+Ucon LB-550X 5+7.5% Shinwasorb-S 100/120
Column Size
3.5 m x 3.0 mm I.D. Stainless steel
Column temp.
80℃
Carrier gas
37 mL/min N2 (Adjust so that the retention time of isoflurane is about 7 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Pyridine , 2-ethyl-2-phenylmalonediamide , Stearylalcohol ,
Applications
Analysis of primidone using Shinwasorb ~Purity test (3) 2-ethyl-2-phenylmalonediamide~
Column
Silicone OV-17 1.5% Shinwasorb-S 100/120
Column Size
1.6 m x 3.2 mm I.D. Glass
Column temp.
195℃
Carrier gas
31 mL/min N2 (Adjust so that the retention time of stearylalcohol is about 10 minutes.)
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Deet, diethyl phthalate in ethanol
Applications
Analysis of deet (Pesticide guidelines) ~Purity test (4) Related substances~
Column
ULBON HR-20M
Column Size
0.2 mm I.D. x 25 m df = 0.1 µmm
Column temp.
170℃
Carrier gas
21 cm/sec He (Adjust so that the retention time of deet is about 10 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL (Split rate: 1:70)

View detail


Graph

GC

Samples
Methanol in ethanol
Applications
Analysis of etidronate disodium using Sunpak-A ~Purity test (4) Methanol~
Column
Sunpak-A 50/80
Column Size
2.1m x3.2 mm I.D. Glass
Column temp.
130℃
Carrier gas
30 mL/min N2 (Adjust so that the retention time of methanol is about 2 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
2-propanol, ethanol in water
Applications
Analysis of sodium aurothiomalate using Sunpak-H ~Purity test (4) Ethanol~
Column
Sunpak-H 80/100
Column Size
3.1m x 3.2 mm I.D. Glass
Column temp.
145℃
Carrier gas
28 mL/min N2 (Adjust so that the retention time of 2-propanol is about 7 minutes.)
Injection port temp.
230°C
Detector temp.
230°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
water, acetonitrile in methanol
Applications
Analysis of panipenem using Sunpak-A ~Water~
Column
Sunpak-A 80/100
Column Size
2.1m x 3.2 mm I.D. Glass
Column temp.
125℃
Carrier gas
29 mL/min He (Adjust so that the retention time of acetonitrile is about 8 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
1.0 μL

View detail


Graph

GC

Samples
Acetone , Ethyl acetate
Applications
Analysis of piperacillin hydrate using Sunpak-H ~(4) Residual solvents~
Column
Sunpak-H 80/100
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
145℃
Carrier gas
33 mL/min N2 (Adjust so that the retention time of ethyl acetate is about 4 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail


Graph

GC

Samples
Cyclohexane , Dicyclohexyl phthalate , Etofenprox
Applications
Analysis of etofenprox using Shinwasorb
Column
Silicone OV-225 2.5% Shinwasorb-U 60/80
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
240℃
Carrier gas
37 mL/min N₂ (Adjust so that the retention time of dicyclohexyl phthalate is about 8 minutes.)
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
Etofenprox in Cyclohexane
Applications
Analysis of etofenprox using Shinwasorb -2- ~Purity (3) Related substances~
Column
Silicone OV-225 2.5% Shinwasorb-U 60/80
Column Size
1.1m x 3.2 mm I.D. Glass
Column temp.
240℃
Carrier gas
37 mL/min N₂
Injection port temp.
250°C
Detector temp.
250°C
Detector
FID
Injection volume
3.0 μL

View detail


Graph

GC

Samples
o-Dichlorobenzene, m-Bromochlorobenzene in acetone
Applications
Analysis of o-dichlorobenzene using Shinwasorb
Column
PEG20M-TPA 2.5% Shinwasorb-U 60/80
Column Size
3.1 m x 3.2 mm I.D. Glass
Column temp.
115℃
Carrier gas
39 mL/min N₂ (Adjust so that the retention time of o-dichlorobenzene is about 11 minutes.)
Injection port temp.
200°C
Detector temp.
200°C
Detector
FID
Injection volume
2.0 μL

View detail