Search result
"Ethyl Mercaptan" Filtered 6 Item.
GC
- Samples
- Carbonyl Sulfide , Hydrogen Sulfide , Carbon Disulfide , Methyl Mercaptan , Ethyl Mercaptan , Dimethyl Sulfide , Benzene(Solvent) , Dimethyl Disulfide
- Applications
- Sulfur Compounds of Air Pollution
- Column
- 1,2,3-TCEP 25% Shimalite 80-100 (AW-DMCS)
- Column Size
- 3.1m×3.2mm
- Column temp.
- 70°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 180°C
- Detector temp.
- 180°C
- Detector
- FPD
View detail
GC
- Samples
- Hydrogen Sulfide , Sulfur Dioxide , Methyl Mercaptan , Ethyl Mercaptan , Dimethyl Sulfide , Carbon Disulfide , Benzene , Dimethyl Disulfide
- Applications
- Air pollution sulfur compound (1)
- Column
- PPE-5rings 10% Shimalite TPA 60-80
- Column Size
- 3.1m×3.2mm
- Column temp.
- 70°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 180°C
- Detector temp.
- 180°C(FPD)
View detail
GC
- Samples
- Methane , Ethane , Hydrogen Sulfide , Cabonyl Sulfide , Propane , Isobutane , Methyl Mercaptane , n-Butane
- Applications
- Sulfur compounds in Lower Hydrocarbon
- Column
- Sunpak-S 80-100
- Column Size
- 2.1m×3.2mm
- Column temp.
- 100°C
- Carrier gas
- N2,50ml/min
- Injection port temp.
- 180°C
- Detector temp.
- 180°C(FPD)
View detail
GC
- Samples
- Carbon Disulfide , Hydrogen Sulfide , Methyl Mercaptan ,Ethyl Mercaptan , DMS , Benzene . DMDS
- Applications
- Air pollution sulfur compound (3)
- Column
- β,β-ODPN 25% Chromosorb W 60-80 (AW-DMCS-ST)
- Column Size
- 3.1m×3.2mm
- Column temp.
- 70°C
- Carrier gas
- N2,50ml/min
- Detector temp.
- 180°C(FPD)
- Injection volume
- -
View detail
GC
- Samples
- Carbonyl sulfide , Hydrogen sulfide , Carbon disulfide , Methyl mercaptan , Ethyl mercaptan , Dimethyl sulfide , Dimethyl disulfide
- Applications
- Analysis of methyl mercaptan, hydrogen sulfide, dimethyl sulfide, and dimethyl disulfide using Shinwasorb
- Column
- β,β’-Oxydipropionitrile 12.5% Shinwasorb-S 60/80 ST
- Column Size
- 3.1 m x 3.2 mm I.D. Glass
- Column temp.
- 70℃
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 180℃
- Detector temp.
- 180℃
- Detector
- FPD (S)
- Injection volume
- 1.0 μL
View detail
GC
- Samples
- Carbon sulfide , Hydrogen sulfide , Carbon disulfide , Methyl mercaptan , Ethyl mercaptane , Dimethyl sulfide
- Applications
- Measurements of offensive odor substances using Shinwasorb -3-
- Column
- 1,2,3-TCEP 25% Shinwasorb-S 80/100 ST
- Column Size
- 3.1 m x 3.2 mm I.D. Glass
- Column temp.
- 70℃
- Carrier gas
- 50 mL/min N₂
- Injection port temp.
- 180℃
- Detector temp.
- 180℃
- Detector
- FPD
- Injection volume
- 1.0 μL
View detail